Plausible but Unimplantable: A Trustworthiness Audit of Shape Completion Foundation Models for Cranial Defect Reconstruction

arXiv cs.CV Medical/Healthcare AI Marek Wodzinski, Kamil Kwarciak, Daria Hemmerling, Mateusz Daniol 2026-10-21

TL;DR - An audit of three pretrained shape-completion models found that none produced implantable reconstructions across 239 cranial defects. The results show that visually plausible skull completion does not imply clinically usable geometry.

  • On 210 artificially carved defects, no audited model improved over simply retaining the patient’s remaining bone; on 29 real cases, only one exceeded chance.
  • Every generated output either intersected existing bone or failed to add meaningful material, making all outputs unimplantable.
  • A task-specific volumetric model was implantable in all but one case, highlighting the gap between general foundation models and specialized systems.
  • Point-cloud outputs at the tested densities could not encode implant thickness, revealing a representation-level limitation beyond individual model quality.